    อุปกรณ์วัด Turbidity (Turbidity Meter with Fast Tracker™ Technology, EPA Compliant)

    HI 98703

    Turbidity Meter
    with Fast Tracker™ Technology, EPA Compliant

     

    • High accuracy at low ranges (below 0.05 NTU)
    • Two, three or four point calibration
    • USB and RS232 PC connectivity
    • Log up to 200 records
    • GLP Features
    • User friendly, backlit  display with guidance codes
    • Battery % on startup
    • Continuous current time on display

     

    The HI 98703 meets and exceeds the requirements of the USEPA Method 180.1 for wastewater and Standard Method 2130 B for drinking water. The instrument has an EPA compliance reading mode which rounds readings to meet EPA reporting requirements. Users will appreciate the accuracy and sensitivity of this instrument, particularly at very low turbidity levels.

    This instrument incorporates complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. The last calibration points,
    time and date can be checked at the touch of a button.

    With it’s logging function, up to 200 measurements along with it’s tagged locations can be stored in internal memory and consulted at any time. Data can be later transferred to a PC via RS232 or USB interface and HANNA HI 92000 software (optional).

    Exclusive Fast Tracker™

    For advanced field applications, the HI 98703 is equipped with Fast Tracker™—Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements.

     

    Order Information:
    HI 98703-01 (115V), HI 98703-02 (230V) and HI 98703-03 (AUS plug) are supplied with sample cuvettes and caps (5), HI 98703-11 calibration cuvettes, HI 93703-58 silicone oil, cuvette cleaning cloth, batteries, AC adapter, instruction manual and rugged carrying case.

     

     Specifications

    Range   0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU
    Range Selection   automatic
    Resolution   0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU
    Accuracy @25°C   ±2% of reading plus 0.02 NTU
    Repeatability   ±1% of reading or 0.02 NTU, whichever is greater
    Stray Light   < 0.02 NTU
    Light Detector   silicon photocell
    Light Source   tungsten filament lamp
    Lamp Life   greater than 100,000 readings
    Method   Ratio Nephelometric Method (90°), ratio of scattered and transmitted light; Adaptation of the USEPA Method 180.1 and Standard Method 2130 B
    Measuring Mode   normal, average, continuous
    Turbidity Standards   <0.1, 15, 100 and 750 NTU
    Calibration   two, three or four-point calibration
    Log Memory   200 records
    PC Connectivity   USB or RS232
    Environment   up to 50°C (122°F); RH max 95% non-condensing
    Power Supply   1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use
    Dimensions / Weight   224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) / 512 g (18 oz.)
  • Requesting a Quotation and Ordering Products

    131 Thailand Science Park. Innovation Cluster 1 Building,

    Klong 1, Klong Luang, Pathumthani 12120 Thailand. 

    Tel.  02-191-3928 Fax 02-191-3929


    For more information please contact 081-847-6768 (Ms. Nareerat) IDLine: @alltestkit

    E-mail : info@highents.com

     

    >>Online Quotation Click here<<