    อุปกรณ์วัด Turbidity (Precision Turbidity Benchtop Meter, EPA Compliant)

    HI 88703

    Precision Turbidity Benchtop Meter, EPA Compliant

     

    • Two measuring ranges: Ratio turbidity, non-ratio turbidity
    • Meets USEPA requirements
    • GLP features        
    • Two, three, four or five point turbidity calibration
    • USB PC connectivity
    • Log up to 200 measurements
    • Contextual help and tutorial mode

     

    The HI 88703 turbidity benchtop meter is specially designed for water quality measurements, providing reliable and accurate readings on low turbidity ranges.

    This instrument has an EPA compliance reading mode which rounds the reading
    to meet EPA reporting requirements. Alternative EBC and Nephelos measuring units are available. Depending on the measured sample and needed accuracy, normal, continuous or signal averaging measurement can be selected.

    A two, three, four or five-point calibration could be performed by using the supplied standards. When user prepared standards are used, calibration points can be modified.
    H 88703 features complete GLP (Good Laboratory Practice) functions that allows traceability of the calibration conditions. The last calibration points, time and date can be checked.

    Up to 200 measurements can be stored in internal memory. Data can be transferred to a PC via optional HI 920013 USB cable and HI 92000 Windows® compatible software.

    Order Information:
    HI 88703-01 (115V) and HI 88703-02 (230V) is supplied with sample cuvettes and caps (5), calibration cuvettes, silicone oil (HI 93703-58), tissue for wiping cuvettes, power cord and instruction manual.

     

     Specifications

    Range Non Ratio Mode 0.00 to 9.99; 10.0 to 40.0 NTU;
    0.0 to 99.9; 100 to 268 Nephelos
    0.00 to 9.80 EBC
      Ratio Mode 0.00 to 9.99; 10.0 to 99.9; 100 to 4000 NTU
    0.0 to 99.9; 100 to 26800 Nephelos
    0.00 to 9.99; 10.0 to 99.9; 100 to 980 EBC
    Resolution Non Ratio Mode 0.01; 0.1 NTU 0.1; 1 Nephelos; 0.01 EBC
      Ratio Mode 0.01; 0.1; 1 NTU 0.1; 1 Nephelos 0.01; 0.1, 1 EBC
    Range Selection   automatic
    Accuracy @25°C   ±2% of reading plus 0.02 NTU (0.15 Nephelos; 0.01 EBC)
    ±5% of reading above 1000 NTU (6700 Nephelos; 245 EBC)
    Repeatability   ±1% of reading or 0.02 NTU (0.15 Nephelos; 0.01 EBC)whichever is greater
    Stray Light   < 0.02 NTU (0.15 Nephelos; 0.01 EBC)
    Light Detector   silicon photocell
    Method   Nephelometric method (90°) or Ratio Nephelometric Method(90° & 180°), Adaptation of the USEPA Method 108.1 and Standard Method 2130 B.
    Measuring Mode   normal, average, continuous
    Turbidity Standards   <0.1, 15, 100, 750 and 2000 NTU
    Calibration   two, three, four or five-point calibration
    Light Source/Life   tungsten filament lamp / greater than 100,000 readings
    Display   40 x 70 mm graphic LCD (64 x 128 pixels) with backlight
    Log Memory   200 records
    PC Interface   USB
    Environment   0°C (32°F) to 50°C (122°F); max 95% RH non-condensing
    Power Supply   230 V/50 Hz or 115 V/60 Hz 20 W; auto-off after 15 minutes of non-use
    Dimensions   230 x 200 x 145 mm (9 x 7.9 x 5.7”)
    Weight   2.5 Kg (88 oz.)

     

  • Requesting a Quotation and Ordering Products

    131 Thailand Science Park. Innovation Cluster 1 Building,

    Klong 1, Klong Luang, Pathumthani 12120 Thailand. 

    Tel.  02-191-3928 Fax 02-191-3929


    For more information please contact 081-847-6768 (Ms. Nareerat) IDLine: @alltestkit

    E-mail : info@highents.com

     

    >>Online Quotation Click here<<