    อุปกรณ์วัด Turbidity (Portable Turbidity Meter with Fast Tracker™ Technology, ISO)

    HI 98713

    Portable Turbidity Meter
    with Fast Tracker™ Technology, ISO

    • Two, three or four point calibration
    • USB and RS 232 PC connectivity
    • Log up to 200 records
    • GLP Features
    • User friendly, backlit LCD display with guidance codes
    • Battery % on startup
    • Continuous current time on display


    The HI 98713 meets and exceeds the requirements of the ISO 7027 for water quality and provides a reliable and accurate readings on low turbidity values. An effective algorithm calculates and converts the detectors output in FNU. Depending on the needed accuracy, normal, continuous, or signal averaging can be selected.

    The optical system—The Infrared Method (ISO 7027), consists of a infrared LED and two detectors (scattered and transmitted light), that assures long term stability and minimizes stray light and color interferences. HI 98713 also compensates for variations in intensity of the LED, minimizing the need for frequent calibration.

    This instrument incorporates complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. The last calibration points, time and date can be checked at the touch of a button.

    With it’s logging function, up to 200 measurements along with it’s tagged locations can be stored in internal memory and consulted at any time. Data can be later transferred to a PC via RS232 or USB interface and HANNA HI 92000 software (optional).

    Exclusive Fast Tracker™

    For advanced field applications, the HI 98713 is equipped with Fast Tracker™—Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements.

     

    Order Information:
    HI 98713-01 (115V) and HI 98713-02 (230V) is supplied with sample cuvettes and caps (5), HI 98713-11 calibration cuvettes, HI 93703-58 silicone oil, cuvette cleaning cloth, batteries, AC adapter, instructions and rugged carrying case.

     

     Specifications

    Range   0.00 to 9.99 FTU
    10.0 to 99.9 FTU
    100 to 1000 FTU
    Range Selection   Automatically
    Resolution   0.01 FTU from 0.00 to 9.99 FTU
    0.1 FTU from 10.0 to 99.9 FTU
    1 FTU from 100 to 1000 FTU
    Accuracy @25°C   ±2% of reading plus 0.1 FTU
    Repeatibility   ±1% of reading or 0.01 FTU, whichever is greater
    Stray Light   < 0.1 FTU
    Light Source / Life   860 nm infrared LED / greater than 100,000 readings
    IR Detector   Silicon Photocell
    Method   Adaptation of ISO 7027, ratio method with 90° and 180° detector.
    Standards   < 0.1, 15, 100 and 750 FTU
    Calibration   Two, three or four point calibration
    Logging Memory   200 records
    PC Connectivity   USB or RS232
    Environment   0 to 50°C (32 to 122°F); max 95% RH non-condensing
    Power Supply   1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use
    Dimensions / Weight   224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) / 512 g (18 oz.)
  • Requesting a Quotation and Ordering Products

    131 Thailand Science Park. Innovation Cluster 1 Building,

    Klong 1, Klong Luang, Pathumthani 12120 Thailand. 

    Tel.  02-191-3928 Fax 02-191-3929


    For more information please contact 081-847-6768 (Ms. Nareerat) IDLine: @alltestkit

    E-mail : info@highents.com

     

    >>Online Quotation Click here<<