    อุปกรณ์วัด Control System (pH Analog Controller with Dual Output and Self-Diagnostic Test)

    HI 8711

    pH Analog Controller with Dual Output and Self-Diagnostic Test

    HI 8711 allows the selection of two set points with two independent outputs for acid and alkaline dosages.

    Each model accepts either a direct input from a pH or ORP electrode or from a transmitter through 4-20 mA input. The instrument also provides ±5V power output and input terminals for amplified electrodes. In addition, you can choose the output configuration for connecting a recorder or a PLC, between 0-20 or 4-20 mA.

    The HI 8711 incorporates adjustable overtime dosing protection from 5 to 60 minutes. If dosing exceeds selected time, the alarm will be triggered and the dosing contact will deactivate. This feature can be activated or deactivated.

    Specifications

    Range  0.00 to 14.00 pH
    Resolution  0.01 pH
    Accuracy
    @ 20 °C  ±0.02 pH (0 to 100 °C)
    ±0.05 pH (-20 to 0 °C)
    ±0.5% (input from transmitter)
    Input  high impedance 1012 Ohm; reference and matching pin inputs are available; 4-20 mA
    Power Output  ±5 Vcc; 150 mA max load for amplified electrodes
    Calibration  offset: ±2 pH with OFFSET trimmer
    slope: 80 to 110% with SLOPE trimmer
    Temperature
    Compensation  fixed or automatic with Pt100, from -20 to 100°C (-4 to 212°F)
    Recorder Output  0-20 mA or 4-20 mA (isolated)
    Setpoint Relay  2, isolated, 2 A, max 240 V, resistive load, 1000000 strokes (not fuse protected)
    Setpoint Range  alk. set: from 0.00 to 14.00 pH
    acid set: from 0.00 to 14.00 pH
    Alarm Relay  1, isolated, 2 A, max 240 V, resistive load, 1000000 strokes (not fuse protected)
    Alarm Range  0.2 to 3.00 pH
    Dosing Control  OFF/AUTO/ON with selection switch
    Over Dosing Control  adjustable, from 5 min to 60 min with knob or disable by wire strap – on rear panel
    Backlight  continuous on
    Power Supply  115 VAC ±10% or 230 VAC ±10%; 50/60 Hz
    Enclosure  flame retardant ABS body and front panel; transparent splash-proof front cover
    Environment  -10 to 50°C (14 to 122°F); RH max 95% non-condensing
    Panel Cutout  141 x 69 mm (5.6 x 2.7″)
    Weight  1 kg (2.2 lb.)

    Accessories

    Solutions
    HI 7004/1L  4.01 pH Buffer Solution
    1 x 1 L bottle
    HI 7007/1L  7.01 pH Buffer Solution
    1 x 1 L bottle
    HI 7010/1L  10.01 pH Buffer Solution
    1 x 1 L bottle
    Accessories
    HI 8427  pH and mV Precision Simulator
    HI 931001  pH and mV Precision Simulator

  • Requesting a Quotation and Ordering Products

    131 Thailand Science Park. Innovation Cluster 1 Building,

    Klong 1, Klong Luang, Pathumthani 12120 Thailand. 

    Tel.  02-191-3928 Fax 02-191-3929


    For more information please contact 081-847-6768 (Ms. Nareerat) IDLine: @alltestkit

    E-mail : info@highents.com

     

    >>Online Quotation Click here<<