    อุปกรณ์วัด Control System (Dissolved Oxygen Controller with Extended Range and Analog Output)

    HI 8410

    Dissolved Oxygen Controller with Extended Range and Analog Output

    The HI 8410 is a panel mounted dissolved oxygen controller that is used to maintain and monitor the concentration of DO in a wide range of industrial process applications. The HI 8410 uses a Galvanic probe that typically requires less maintenance than a Polarographic style making it ideal for long term monitoring.

    The set point for controlling the activation of a relay is adjusted manually by the user. An alarm relay is also manually adjustable and is based upon a tolerance from the programmed set point. Calibration is single set point and can be done in zero oxygen solution.

    The dosage mode: automatic, continuous ON or OFF and over dosage control by setting the overtime dosage trimmer. If the dosing relay remains continuously activated for more than the selected dosing time, the alarm relay is activated, the alarm LED will start blinking and the dosing relay will be deactivated. A jumper found on the rear panel can disable the “over time dosage” function.

    “Automatic/Off/manual” dosing selection switch and LED on the front panel. In Automatic mode all the relays are controlled based on the measurement set point and alarm values. In OFF mode the dosing and alarm relays are always deactivated. The dosing LED is OFF (as relay status) and the ALARM LED is in accordance with the instrument set point, input reading, and ALARM. In ON (Manual) mode the dosing relay is always on. The alarm relay is still enabled. If an alarm occurs the dosing relay remains activated. If the over dose time exceeds the setting during manual mode, the alarm relay remains activated.

    The D.O. probe is provided with a membrane covering the galvanic sensor and a built-in thermistor for temperature measurement and compensation.

    Other features include: recorder output in 0-20 mA or 4-20 mA configuration, LED indicators which identify whether the controller is in operation mode or setup selection mode, overtime control function and hysteresis setting.

    Specifications

    Range  0.0 to 50.0 mg/L (ppm) O2
    0 to 600 % O2
    -5.0 to 50.0 °C
    Resolution  0.1 mg/L or 1% (O2) / 0.1 °C
    Accuracy @ 20°C  ±1% of reading (O2) / ±0.2 °C
    Calibration  manual, one point, in saturated air
    Temperature
    Compensation  automatic, from -5 to 50°C (23 to 122 °F)
    Salinity Compensation  0 to 51 g/L (resolution 1 g/L)
    Probe (not included)  HI 76410/4 with 4 m (13.1’) cable or HI 76410/10 with 10 m (32.8’) cable
    Recorder Output  0 to 20 mA or 4 to 20 mA (isolated)
    Setpoint and
    Alarm Relay  1, isolated, 2A, max. 240V, resistive load, 1,000,000 strokes
    Setpoint Range  1 to 600 % O2; 0.1 to 50.0 mg/L (ppm) O2
    Alarm Range  1.0 to 5.0 mg/L (ppm) O2
    Hysteresis Range  0.5 to 2.4 mg/L (ppm) O2
    Dosing Control  OFF/AUTO/ON with selection switch
    Over Dosing Control  adjustable, from 5 min to 60 min with knob or disable by wire strap – on rear panel
    Backlight  continuous on
    Power Supply  115 VAC ±10% or 230 VAC ±10%; 50/60 Hz
    Enclosure  flame retardant ABS body and front panel; transparent splash-proof front cover
    Environment  -10 to 50°C (14 to 122°F); RH max 95% non-conndensing
    Panel Cutout  141 x 69 mm (5.6 x 2.7’’)
    Weight  1 kg (2.2 lb.)

  • Requesting a Quotation and Ordering Products

    131 Thailand Science Park. Innovation Cluster 1 Building,

    Klong 1, Klong Luang, Pathumthani 12120 Thailand. 

    Tel.  02-191-3928 Fax 02-191-3929


    For more information please contact 081-847-6768 (Ms. Nareerat) IDLine: @alltestkit

    E-mail : info@highents.com

     

    >>Online Quotation Click here<<